Note: a minimum order consists of 20 units;
product shipment is approximately three days after receipt of order.
 |
 |
 |
| c\temp |
|
Data Sheet |
 |
| Temperature Range |
|
c\temp: -29°C [-20°F] to 72°C [160°F] Red
Case
c\temp dsc: -40°C [-40°F] to 72°C
[160°F] Gray Case
c\temp dlt: -80°C [-113°F] to 30°C [86°F] Blue
Case |
 |
| Temperature Accuracy |
|
0.2°C [0.4°F] at 20°C [68°F]
0.5°C [1.0°F] at extremes |
 |
| Memory EPROM |
|
8K storage |
 |
| Measurement Interval (pre-programmed by factory) |
|
5
days = 58 sec.
7 days = 1 min. 20 sec.
10 days = 1 min. 54 sec.
30 days = 5 min. 38 sec.
60 days = 11 min. 16 sec. |
 |
| Computer Interface |
|
Serial port RS232 serial cable |
 |
| Indicators |
|
Two: Start/Stop and Alarm |
 |
| High and Low Alarm Limits |
|
LED indicator alarm set by temperature and time |
 |
| Internal Software Features |
|
Times, start time, interval, user description, alarm values, recording
status, measurement data, software version, serial number,
model type. |
 |
| Power Source |
|
c\temp - 3.0v Lithium battery (1 year shelf life)
c\temp dsc - 3.6v Lithium battery (1 year shelf life) |
 |
| Size |
|
Centimeters: 9(L) x 5(W) x 1.6(H) Inches: 3.54"(L) x 1.97"(W)
x 0.63"(H) |
 |
| Weight |
|
52.5 grams [1.85 ounces] |
 |
| CE Certified |
|
Yes |
 |
| c\temp Software features |
|
Report Summary & Statistics
Mean Kinetic Temperature Calculation
File Manager
Note attachment
Zoom
Ability to look at recorded temperature points and date / time stamp
Elapsed time view or date / time recorded view
Celsius, Fahrenheit or Kelvin display
3D view option
Can open up to 16 files at one time
Export data to CSV or TXT file
Export graph as a bitmap or metafile |
 |